Pemanasan Tabung
Tube Warm-Up
Perhatian!
Prosedur PEMANASAN WAJIB DILAKUKAN apabila tabung X-Ray TIDAK DIGUNAKAN LEBIH DARI 24 JAM
The warm-up procedure is MANDATORY if the X-Ray tube HAS NOT BEEN USED FOR MORE THAN 24 HOURS
Prosedur PEMANASANwarm-up procedure
- Kolimator harus ditutup sepenuhnya dan diarahkan ke area yang kosong.The collimator must be fully closed and aimed at an empty area.
- Pengguna harus menggunakan APD.Users must wear Personal Protective Equipment.
- Nyalakan unit.Turn on the unit.
- Lakukan prosedur pemanasan tabung sesuai dengan tabel berikut:Follow the tube warm-up procedure according to the table below:
| Phase | kVp | mA | ms | Total Exposure | Interval |
|---|---|---|---|---|---|
| 1 | 60 | 50 | 500 | 1 | 20 seconds |
| 2 | 60 | 50 | 500 | 1 | 20 seconds |
| 3 | 80 | 50 | 100 | 1 | 20 seconds |
| 4 | 80 | 50 | 100 | 1 | 20 seconds |
| 5 | 100 | 25 | 500 | 1 | 20 seconds |
| 6 | 100 | 25 | 500 | 1 | 20 seconds |