Prosedur Forming
Forming Procedure
Perhatian!
Prosedur FORMING WAJIB DILAKUKAN apabila tabung X-Ray TIDAK DIGUNAKAN LEBIH DARI 30 HARI
The forming procedure is MANDATORY if the X-Ray tube HAS NOT BEEN USED FOR MORE THAN 30 DAYS
Prosedur FORMINGforming procedure
- Kolimator harus ditutup sepenuhnya dan diarahkan ke area yang kosong.The collimator must be fully closed and aimed at an empty area.
- Pengguna harus menggunakan APD.Users must wear Personal Protective Equipment.
- Nyalakan unit.Turn on the unit.
- Lakukan prosedur forming tabung sesuai dengan tabel berikut:Follow the tube forming procedure according to the table below:
| Phase | kVp | mA | ms | Total Exposure | Interval |
|---|---|---|---|---|---|
| 1 | 70 | 25 | 500 | 2 | 60 seconds |
| 2 | 75 | 25 | 500 | 2 | 60 seconds |
| 3 | 80 | 25 | 500 | 2 | 60 seconds |
| 4 | 85 | 25 | 500 | 2 | 60 seconds |
| 5 | 90 | 25 | 500 | 2 | 60 seconds |
| 6 | 95 | 25 | 500 | 2 | 60 seconds |
| 7 | 100 | 25 | 500 | 2 | 60 seconds |
| 8 | 105 | 25 | 500 | 2 | 60 seconds |
| 9 | 110 | 25 | 500 | 2 | 60 seconds |